DISCLAIMER: The Graph Chatbot is not programmed to provide explicit or categorical answers to your questions. Rather, it transforms your questions into API requests that are distributed across the various IT services officially administered by EPFL. Its purpose is solely to collect and recommend relevant references to content that you can explore to help you answer your questions.
We use magnetic microbeads, which are magnetically self-assembled in chains in a microfluidic chip, as reaction substrates to implement two different sandwich immunoassay protocols for the detection of mouse monoclonal target antibodies. The magnetic chains form when the chip is placed in a magnetic field, and are geometrically trapped and accurately positioned in a microchannel with periodically enlarged cross-sections. In the first immunoassay protocol, capture and target antibodies are incubated off-chip, while exposure to the detection antibody is performed on-chip. In the second protocol, the complete immunoassay is fully executed on-chip. In the ‘off-chip incubation–on-chip detection’ protocol, antibodies can be detected down to a concentration of 50 ng/mL in a total assay time of 120 min, while consuming 1.5 mL of target antibody solution. Using the full on-chip protocol, our system is able to detect target antibodies in the range of a few ng/mL in 30 min, using only a few tens of nanoliters of target antibody solution and reagents. The ‘off-chip incubation–on-chip detection’ protocol is also applied for dosing antibodies obtained from the supernatant of a cell culture medium.
Christoph Merten, Leonie Kolmar, Xiaoli Ma, Hongxing Hu