DISCLAIMER: The Graph Chatbot is not programmed to provide explicit or categorical answers to your questions. Rather, it transforms your questions into API requests that are distributed across the various IT services officially administered by EPFL. Its purpose is solely to collect and recommend relevant references to content that you can explore to help you answer your questions.
This paper is focused on modeling and performance investigation of insulin injection pens used by diabetic persons. A smart pen cap with two electrodes is used as a capacitive sensor for insulin dose detection and is presented in details. To faithfully analyze this medical device, both an analytical model and a fully parametrized three-dimensional physical model in ANSYS have been developed. The electrode capacitance is evaluated (by both simulations and analytical model), for the smart pen cap alone as well as for different insulin fill states in the insulin injection pen + smart pen cap complete system.
Martinus Adela Maria Gijs, Martin Rainer Wohlwend
Silvestro Micera, Stéphanie Lacour, Valentina Marie Paggi