We report on the design and commissioning of a new spectrometer for muon-spin relaxation/rotation studies installed at the Swiss Muon Source (SμS) of the Paul Scherrer Institute (PSI, Switzerland). This new instrument is essentially a new design and replaces the old general-purpose surface-muon (GPS) instrument that has been for long the workhorse of the μSR user facility at PSI. By making use of muon and positron detectors made of plastic scintillators read out by silicon photomultipliers, a time resolution of the complete instrument of about 160 ps (standard deviation) could be achieved. In addition, the absence of light guides, which are needed in traditionally built μSR instrument to deliver the scintillation light to photomultiplier tubes located outside magnetic fields applied, allowed us to design a compact instrument with a detector set covering an increased solid angle compared with the old GPS.
Jian Wang, Matthias Finger, Qian Wang, Yiming Li, João Miguel das Neves Duarte, Matthias Wolf, Varun Sharma, Yi Zhang, Tian Cheng, Yixing Chen, Alexis Kalogeropoulos, Ioannis Papadopoulos, Hua Zhang, Siyuan Wang, Xin Chen, Michele Bianco, Sebastiana Gianì, Sun Hee Kim, Rakesh Chawla, Jan Steggemann, Konstantin Androsov, Anna Mascellani, Federica Legger, Matteo Galli, Gabriele Grosso