DISCLAIMER: The Graph Chatbot is not programmed to provide explicit or categorical answers to your questions. Rather, it transforms your questions into API requests that are distributed across the various IT services officially administered by EPFL. Its purpose is solely to collect and recommend relevant references to content that you can explore to help you answer your questions.
In the present work we intend to assess the readout feasibility of Avalanche Photodiode (APD) detectors via long signal lines for the development of a combined small-animal PET-MR prototype based on the ClearPEM technology. The detection performance of a LYSO-APD module was evaluated reading out the APD charge signals to the front-end ASIC via an 80 mm length flexible flat-cable (FFC). Experimental results showed a time resolution of 5.06 ns for the detector module in double-readout mode, with a nonsignificant degradation of 8.4% with the introduction of the FFC. The energy resolution of the system was not degradated by the FFC. (C) 2012 Elsevier B.V. All rights reserved.