Mediaspace scheduled maintenance: Aug 25, 2026 07:00 - 12:00 AM. During this time, videos will be temporarily unavailable. Check status updates.
This paper systematically demonstrates the feasibility of wirelessly monitoring dopamine concentration in the brain with an implantable biosensor. The biosensor was realized using microwires, and then, the dopamine concentration was measured in-vitro ranging from 0.3 μM to 2 μM, corresponding to the physio-pathological concentration range in human brain. The obtained results were used to design and optimise a full-custom CMOS sensor interface for in-vivo dopamine monitoring. The key component of this interface is a potentiostat with a maximum power consumption of 10.24 μW in a 10 kHz sampling frequency. The CMOS interface automatically subtracts the background current up to 2.34 μA. The obtained sensitivity in dopamine detection has been evaluated in 150 μA/μM, with a Limit of Detection (LoD) of 33 nM, thus being suitable for dopamine monitoring in human brain.
Patric Hagmann, Elda Fischi Gomez, Oscar Esteban Rucavado Leandro, Jonas Richiardi, Thomas Sanchez
Dimitri Nestor Alice Van De Ville, Olaf Blanke, Giulio Rognini, Thomas William Arthur Bolton, Nathan Quentin Faivre, Eva Blondiaux, Herberto Dhanis Pedro De Barros Camacho