DISCLAIMER: The Graph Chatbot is not programmed to provide explicit or categorical answers to your questions. Rather, it transforms your questions into API requests that are distributed across the various IT services officially administered by EPFL. Its purpose is solely to collect and recommend relevant references to content that you can explore to help you answer your questions.
A different approach in microcolumn design is presented, aiming at a large number of pixels at minimal probe size for the deflected beam. An optimization routine resulted in a seven times magnifying column featuring a more than 7×7 mm2 scan field at 40 mm working distance. Simulations for 1 keV electrons from a field emission source predict an increase in beam size from 85 nm on axis up to only about 200 nm for a beam deflected 3 mm off axis. Within a 1 mm scan field this microcolumn could address over 100 Mpixels of less than 100 nm in size. Tests of this design using the 130 nm electron probe of a scanning electron microscope as the electron source resulted in a beam size of ~930 nm on axis up to ~1000 nm for a beam deflected 3 mm off axis.