Ask any question about EPFL courses, lectures, exercises, research, news, etc. or try the example questions below.
DISCLAIMER: The Graph Chatbot is not programmed to provide explicit or categorical answers to your questions. Rather, it transforms your questions into API requests that are distributed across the various IT services officially administered by EPFL. Its purpose is solely to collect and recommend relevant references to content that you can explore to help you answer your questions.
Venous cannula orifice obstruction is an underestimated problem during augmented cardiopulmonary bypass (CPB), which can potentially be reduced with redesigned, virtually wall-less cannula designs versus traditional percutaneous control venous cannulas. A bench model, allowing for simulation of the vena cava with various affluent orifices, venous collapse and a worst case scenario with regard to cannula position, was developed. Flow (Q) was measured sequentially for right atrial + hepatic + renal + iliac drainage scenarios, using a centrifugal pump and an experimental bench set-up (afterload 60 mmHg). At 1500, 2000 and 2500 RPM and atrial position, the Q values were 3.4, 6.03 and 8.01 versus 0.77*, 0.43* and 0.58* l/min: p
Yves Perriard, Yoan René Cyrille Civet, Armando Matthieu Walter, Amine Benouhiba, Francesco Clavica, Silje Ekroll Jahren