DISCLAIMER: The Graph Chatbot is not programmed to provide explicit or categorical answers to your questions. Rather, it transforms your questions into API requests that are distributed across the various IT services officially administered by EPFL. Its purpose is solely to collect and recommend relevant references to content that you can explore to help you answer your questions.
In this work, we showed the realization of a fully- implantable device that integrates a microfabricated sensing platform, a coil for power and data transmission and integrated circuits. We described a device intended to test the biocompati- bility of the materials used for the microfabrication. Therefore, electronics measurements for data communication and remote powering will be reported in another article [1]. To ensure biocompatibility an epoxy enhanced polyurethane membrane was used to cover the device. We proved through an in-vitro characterization that the membrane was capable to retain enzyme activity up to 35 days. After 30 days of implant in mice, in-vivo experiments proved that the membrane promotes the integration of the sensor with the surrounding tissue, as demonstrated by the low inflammation level at the implant site
Stefania Maria Aliki Konstantinidi