DISCLAIMER: The Graph Chatbot is not programmed to provide explicit or categorical answers to your questions. Rather, it transforms your questions into API requests that are distributed across the various IT services officially administered by EPFL. Its purpose is solely to collect and recommend relevant references to content that you can explore to help you answer your questions.
We present fast high-roughness and non-contact surface measurements by digital holographic microscopy (DHM). By using single- and dual-wavelength operation modes, coupled with advanced image stitching and non-measured points management methods, the technique enables two-dimensional roughness measurements up to the micrometer (N6). The sample is mechanically scanned over a surface up to 5 × 0.3 mm2 with 17 holograms each acquired in less than 500 µs, the corresponding phase images stitched together by software, and therefore providing multiple profiles measurement in the ISO definition in less than 30 s. The approach is validated by inspection of several different roughness standards and our technique is demonstrated to be in agreement with two existing well-known techniques in the field.