DISCLAIMER: The Graph Chatbot is not programmed to provide explicit or categorical answers to your questions. Rather, it transforms your questions into API requests that are distributed across the various IT services officially administered by EPFL. Its purpose is solely to collect and recommend relevant references to content that you can explore to help you answer your questions.
An original electrical AC loss experimental set-up is presented. For loss comparison purpose, Ic and exponent n are measured with a DC magnetic field between 2 and 400 mT applied perpendicularly to the axis of the sample using a specially designed electromagnet. A computer controlled stepper motor allows a the field angle to vary from parallel to perpendicular with respect to the tape surface. Using a triangular input current, voltage is measured and averaged, and then fitted to a power law in the form V0*(I/Ic)^n. The measurements are repeated to give the dependencies of Ic and n on the magnetic field amplitude and angle.