DISCLAIMER: The Graph Chatbot is not programmed to provide explicit or categorical answers to your questions. Rather, it transforms your questions into API requests that are distributed across the various IT services officially administered by EPFL. Its purpose is solely to collect and recommend relevant references to content that you can explore to help you answer your questions.
This paper presents an energy-efficient analog front-end circuit to monitor the pulmonary artery pressure (PAP) with piezoresistive pressure sensors. A low-power capacitively-coupled instrumentation amplifier (CCIA) is developed in order to boost the amplitude of the bridge sensor to the input range of the analog-to-digital converter (ADC). The 1/f noise and input-referred offset of the amplifier are eliminated by the chopping technique. Simulation results show that this CCIA with chopping technique achieves an input-referred noise density of 38.7 nV/p Hz and an input-referred offset voltage of 4 mu V with a current consumption of 4.28 mu A at 1.8 V voltage supply. This gives a noise-efficiency factor (NEF) of 3.1. Dual supply voltages are used to decrease the power consumption of the bridge sensor. The pressure range is selected to be between 0 and 125 mmHg and the bandwidth of the interface is set from DC to 400 Hz with an RC low pass filter (LPF). This low-noise, low-offset analog front-end circuit achieves an accuracy of 0.125 mmHg.