DISCLAIMER: The Graph Chatbot is not programmed to provide explicit or categorical answers to your questions. Rather, it transforms your questions into API requests that are distributed across the various IT services officially administered by EPFL. Its purpose is solely to collect and recommend relevant references to content that you can explore to help you answer your questions.
Existing clinical methods for bacteria detection lack speed, sensitivity, and, importantly, point-of-care (PoC) applicability. Thus, finding ways to push the sensitivity of clinical PoC biosensing technologies is crucial. Here we report a portable PoC device based on lens-free interferometric microscopy (LIM). The device employs high performance nanoplasmonics and custom bioprinted microarrays and is capable of direct label-free bacteria (E. coli) quantification. With only one-step sample handling we offer a sample-to-data turnaround time of 40 min. Our technology features detection sensitivity of a single bacterial cell both in buffer and in diluted blood plasma and is intrinsically limited by the number of cells present in the detection volume. When employed in a hospital setting, the device has enabled accurate categorization of sepsis patients (infectious SIRS) from control groups (healthy individuals and noninfectious SIRS patients) without false positives/negatives. User-friendly on-site bacterial clinical diagnosis can thus become a reality.
Andrea Rinaldo, Cristiano Trevisin
Camille Véronique Bernadette Goemans, Christian Eugen Zimmerli