DISCLAIMER: The Graph Chatbot is not programmed to provide explicit or categorical answers to your questions. Rather, it transforms your questions into API requests that are distributed across the various IT services officially administered by EPFL. Its purpose is solely to collect and recommend relevant references to content that you can explore to help you answer your questions.
An interferometric method to record quantitative X-ray phase contrast images has been developed that can be used at polychromatic and incoherent X-ray sources such as laboratory tubes. With respect to previously presented results, in this work we report on recent developments and results that have been achieved in view of potential future applications such as in medicine or biology. In particular, due to improvements in the fabrication process large area diffraction gratings with high aspect ratio were achieved. Thereby, the field of view of the interterometer has been drastically increased to 64 x 64 mm(2) and the design value of the photon energy for the gratings could be increased up to 28 keV. Moreover, the use of a Medipix2 single photon-counting pixel detector shows a considerable improvement in image quality and sensitivity over the integrating detector used so far.