DISCLAIMER: The Graph Chatbot is not programmed to provide explicit or categorical answers to your questions. Rather, it transforms your questions into API requests that are distributed across the various IT services officially administered by EPFL. Its purpose is solely to collect and recommend relevant references to content that you can explore to help you answer your questions.
We present a fully integrated detector suitable for active tracking of interventional devices in MR-guided interventions. The single-chip microsystem consists of a detection coil, a tuning capacitor, an intermediate frequency downconversion receiver, and a phase-locked-loop-based frequency synthesizer. Thanks to the integrated mixer, the chip output stage delivers an analog frequency-downconverted NMR signal in the frequency range from 0 to 200 kHz. The microchip, realized in a standard complementary metal oxide semiconductor technology, has a size of 1 x 2 x 0.74 mm3 and operates at a frequency of 63 MHz (i.e., in 1.5 T clinical scanners). Tests in a standard clinical scanner demonstrate the compatibility of the complementary metal oxide semiconductor microchip with clinical MRI systems. Using a solid sample of cis-polyisoprene having a size of 1 x 1.9 x 0.8 mm3 as internal signal source, the detector achieves a three-dimensional isotropic spatial resolution of 0.15 mm in a measuring time of 100 ms. Magn Reson Med, 2011. (C) 2011 Wiley-Liss, Inc.
Edoardo Charbon, Claudio Bruschini, Ekin Kizilkan, Utku Karaca
Pasquale Scarlino, Franco De Palma, Fabian Oppliger