DISCLAIMER: The Graph Chatbot is not programmed to provide explicit or categorical answers to your questions. Rather, it transforms your questions into API requests that are distributed across the various IT services officially administered by EPFL. Its purpose is solely to collect and recommend relevant references to content that you can explore to help you answer your questions.
This letter presents the first neural front-end specifically targeting intravascular recording with direct frequency-domain encoding for body-coupled communication. It employs a differential amplifier and voltage-controlled oscillator (VCO)-based amplitude-to-frequency encoding. Fabricated in TSMC 180 nm CMOS technology, the proposed architecture integrates a fully differential capacitive-feedback low-noise amplifier with a current-starved ring VCO to directly translate microvolt-level neural signals into frequency variations. The design occupies 0.0025 mm2 and operates from externally supplied ±0.6V DC rails, intended to be delivered through the catheter in the target application, without any on-chip negative-voltage generator. Measured results demonstrate linear amplitude-to-frequency conversion up to 2mVpp input with a minimum detectable signal of 20 μVpp. The measured bandwidth is 6 Hz–6.7kHz. The total power consumption is 11 μW under PCB loading conditions. The integrated equivalent input-referred noise is 5.55 μVrms over the spike band (300 Hz–6 kHz). By prioritizing extreme miniaturization and frequency-domain robustness, the proposed design enables neural interfaces compatible with catheter-based deployment and body-coupled transmission.