DISCLAIMER: The Graph Chatbot is not programmed to provide explicit or categorical answers to your questions. Rather, it transforms your questions into API requests that are distributed across the various IT services officially administered by EPFL. Its purpose is solely to collect and recommend relevant references to content that you can explore to help you answer your questions.
Elevated body temperature (fever) can be a common symptom of a medical condition, such as a viral or bacterial infection, including SARS-CoV-2 or influenza. Non-contact infrared thermometers are able to measure forehead temperature in a timely manner and were used to perform a fast fever screening in a population. However, forehead temperature measurements differ greatly from basal body temperatures, and are the target of massive perturbations from the environment. Here we gathered a dataset of N=18024 measurements using the same precision infrared sensor in different locations while tracking both outside temperature, room temperature, time of measurement, and identity. Herein, we propose a method able to extract and remove the influence of external perturbations, and set the threshold for fever detection to 37.38 degrees C, after calibration and temperature correction. This method can help manufacturers and decision-makers to build and use more accurate tools so as to maximize both sensitivity and specificity of the screening protocol.