DISCLAIMER: The Graph Chatbot is not programmed to provide explicit or categorical answers to your questions. Rather, it transforms your questions into API requests that are distributed across the various IT services officially administered by EPFL. Its purpose is solely to collect and recommend relevant references to content that you can explore to help you answer your questions.
Sub-ppm hydrogen chloride (HCl) and water vapour (H2O) monitoring using photoacoustic spectroscopy in optical fibre manufacturing is reported. The development and performance of a sensor based on an acoustic resonant configuration is described, and on-site measurements are presented. Two DFB lasers emitting in the 1370 nm and 1740 nm range were used for the detection of H2O and HCl, respectively. A detection limit (defined for a SNR=3) of 60 ppb for HCl and 40 ppb for H2O was achieved. Contamination sources of the carrier gas used for the fibre preform manufacturing are identified and discussed. © Springer-Verlag 2006.