DISCLAIMER: The Graph Chatbot is not programmed to provide explicit or categorical answers to your questions. Rather, it transforms your questions into API requests that are distributed across the various IT services officially administered by EPFL. Its purpose is solely to collect and recommend relevant references to content that you can explore to help you answer your questions.
We report on the design and operation of a world-to-chip microfluidic interface and experimental setup for fluidic micro- and nano-electromechanical systems. The central component of the interface is an engineered polyether ether ketone connector that brings fluid samples from a commercial syringe pump to the chip with the help of o-rings. In addition to that, the connector serves as an on-chip vacuum chamber. To confirm the adequate operation of our interface, we use complex microfluidic devices that were previously fabricated, suspended microchannel resonators, and demonstrate a fast exchange between fluids (on the scale of 130 s from isopropyl alcohol to water), in-vacuum operation of the devices (intrinsic damping regime), and accurate temperature control of the chip at different set points.
Mark Pauly, Quentin Christian Becker
Olga Fink, Keivan Faghih Niresi