DISCLAIMER: The Graph Chatbot is not programmed to provide explicit or categorical answers to your questions. Rather, it transforms your questions into API requests that are distributed across the various IT services officially administered by EPFL. Its purpose is solely to collect and recommend relevant references to content that you can explore to help you answer your questions.
This paper presents an energy-efficient, duty-cycled, and spinning excitation bridge-to-digital converter (BDC) suitable for measuring the pulmonary artery pressure of heart failure patients with an implantable system. The duty-cycled bridge uses resistances of 6.2 kΩ and, with a supply of 1.2 V, consumes 0.4 nJ excitation energy. A novel spinning method is applied to the bridge and the capacitive DAC simultaneously in such a way to achieve an offset-independent digital output and to eliminate the need for complex instrumentation amplifiers with offset-reduction techniques or calibration. The SAR ADC fabricated in 0.18-µm CMOS consumes 19 nW at 1.2 V. With a sampling rate of 1 kS/ s , the converter achieves the ENOB of 9.2 bits.