Ask any question about EPFL courses, lectures, exercises, research, news, etc. or try the example questions below.
DISCLAIMER: The Graph Chatbot is not programmed to provide explicit or categorical answers to your questions. Rather, it transforms your questions into API requests that are distributed across the various IT services officially administered by EPFL. Its purpose is solely to collect and recommend relevant references to content that you can explore to help you answer your questions.
A silicon-based temperature stabilized axial Hall probe with an absolute accuracy of +/-40 ppm in the full range between 0 and 6 T is presented. The absolute calibration of the probe is performed against a nuclear magnetic resonance (NMR) probe. The application of the probe to a magnetic field profile of a 5 T superconducting magnet system, with field gradients as high as 30 T/m, reaching an overall accuracy of better than +/-100 ppm, is demonstrated. In addition to the high absolute accuracy, this Hall probe allows high spatial resolution measurements of inhomogeneous fields in configurations where present NMR probes are not usable. (C) 1999 American Institute of Physics. [S0034-6748(99)01506-3].